Command Reference
R&S
®
ZNB/ZNBT
1285User Manual 1173.9163.02 ─ 55
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection:
TCOefficient:CALibration
Usage: Query only
Manual operation: See "Frequency / Transm. Coefficients: Insert, Delete, Delete
All" on page 493
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:DEFine<ListNo>
<Frequency>, <TransCoeff>
Adds a new frequency and transmission coefficient to the end of the power loss list.
The query returns the frequency and transmission coefficient no. <ListNo>.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the trans-
mission coefficient settings are valid for all channels.
<PhyPt> Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
<ListNo> Number of point in the list. This suffix is only used for queries.
Parameters:
<Frequency> Stimulus frequency value. If several points with identical fre-
quencies are added, the analyzer automatically ensures a fre-
quency spacing of 1 Hz.
Range: Stimulus values outside the frequency range of the
analyzer are allowed.
*RST: n/a
Default unit: Hz
<TransCoeff> Transmission coefficient
Range: -300 dB to +200 dB
*RST: n/a
Default unit: dB
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection:
TCOefficient:CALibration
Manual operation: See "Frequency / Transm. Coefficients: Insert, Delete, Delete
All" on page 493
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:DELete<ListNo>:ALL
Clears the power loss list.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the trans-
mission coefficient settings are valid for all channels.
<PhyPt> Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
SCPI Command Reference