Command Reference
R&S
®
ZNB/ZNBT
1329User Manual 1173.9163.02 ─ 55
Parameters:
<Boolean> ON - display conductances
OFF - display capacitaces
Manual operation: See "Conductance in Embedding Networks" on page 703
SYSTem:DISPlay:DIALogs:SETup:MCAL[:STATe] <Boolean>
Enables/disables multiple calibrations in the calibration wizard.
Parameters:
<Boolean>
Manual operation: See "Multiple Cal in Calibration Wizard" on page 701
SYSTem:DISPlay:TRACes:CCOunt
<ColorCount>
Defines the maximum number of trace colors (trace properties for the user defined
color scheme defined using DISPlay:CMAP<DispEl>:RGB.
Parameters:
<ColorCount> Range: 1 to 100
Manual operation: See "Number of Trace Colors" on page 703
SYSTem:DISPlay:UPDate <Activate>
Switches the display on or off while the analyzer is in the remote state. The command
has no effect while the analyzer is in the local operating state.
Tip: Switching off the display speeds up the measurement. This command may have
an impact on the update of trace and channel settings; see SYSTem:SETTings:
UPDate.
Parameters:
<Activate> ON | OFF - switch the display on or off. If the display is switched
on, the analyzer shows the diagrams and traces like in manual
control.
ONCE - switch the display on and show the current trace. This
parameter can be used for occasional checks of the measure-
ment results or settings. The measurement is continued, how-
ever, the measurement results are not updated. Compared to
the ON setting, ONCE does not slow down the measurement
speed.
Example:
SYST:DISP:UPD ON
Switch the display on to view the traces and diagrams.
SYSTem:ERRor[:NEXT]?
Queries and at the same time deletes the oldest entry in the error queue. Operation is
identical to that of STATus:QUEue[:NEXT]?
SCPI Command Reference