Command Reference
R&S
®
ZNB/ZNBT
1337User Manual 1173.9163.02 ─ 55
SYSTem:PRESet:STARt <Boolean>
Defines whether the analyzer firmware always starts with the configured system preset
(factory or user defined) or with the previous state. This is a global setting.
Parameters:
<Boolean> ON
The FW always starts with the configured system preset.
See SYSTem:PRESet:USER:NAME and SYSTem:PRESet:
USER[:STATe].
OFF
The analyzer firmware recalls the previous state, i.e. the settings
that were persisted during the previous shutdown.
Manual operation: See "Start in Preset" on page 700
SYSTem:PRESet:USER[:STATe]
<Boolean>
Determines whether a factory or a user-defined preset is performed.
This setting applies to the analyzer GUI and to the remote commands *RST and
SYSTem:PRESet[:DUMMy], if SYSTem:PRESet:REMote[:STATe] is set to ON.
Parameters:
<Boolean> OFF (0)
A factory preset is performed.
ON (1)
A user-defined preset is performed, if the setup file selected
using SYSTem:PRESet:USER:NAME is available and valid. Oth-
erwise, a factory preset is performed.
Example:
SYST:PRES:USER:NAME 'C:\Users\Public\Documents
\Rohde-Schwarz\Vna\RecallSets\Setup_2.znx'
Select a setup file for the user-defined preset.
SYST:PRES:USER ON
Enable the user-defined preset.
SYSTem:PRESet:REMote ON
Align *RST (and SYSTem:PRESet) to the user-defined preset.
*RST
Perform the user-defined preset.
Manual operation: See "Preset Configuration" on page 699
SYSTem:PRESet:USER:CAL <PresetUserCal>
Selects a calibration from the calibration pool that shall be restored during a user-
defined preset.
The corresponding cal group file (<PresetUserCal>.cal) must be available in the
cal pool directory
C:\Users\Public\Documents\Rohde-Schwarz\Vna\Calibration\Data\.