Command Reference
R&S
®
ZNB/ZNBT
1354User Manual 1173.9163.02 ─ 55
CALCulate<Chn>:LIMit:CONTrol:DOMain <SweepType>
Deletes the existing limit line and (re-)defines the physical units of the stimulus values
of the limit line. The units of the response values and the scaling of the y-axis can be
defined via CALCulate<Ch>:LIMit:RDOMain:...
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Setting parameters:
<SweepType> FLIN | FLOG | FSEG | FSINgle | TLIN | TLOG | PLIN | PLOG |
PSINgle
Keywords for the units of the stimulus values. The selected unit
must be compatible with the sweep type ([SENSe<Ch>:
]SWEep:TYPE); otherwise the limit line can not be displayed
and no limit check is possible.
The parameters form three groups: FLIN, FLOG, FSEG, and
FSINgle select frequency units for the limit line. TLIN and TLOG
select time units, PLIN, PLOG and PSINgle select power units.
*RST: FLIN
Default unit: Hz (for FLIN, FLOG, FSEG, FSINGle); s (for TLIN,
TLOG), dBm (for PLIN, PLOG, PSINgle).
Example:
SWE:TYPE POW
Select a power sweep.
CALC:LIM:CONT:DOM PLIN
Delete all existing limit line segments and select level units for
the limit line of the active trace.
CALC:LIM:CONT -20, -10
Define a limit line segment in the stimulus range between -20
dBm and -10 dBm.
Usage: Setting only
CALCulate<Chn>:LIMit:RDOMain:COMPlex <UnitRef>
Deletes the existing limit line and (re-)defines the physical units of the response values
of the limit line. The units of the stimulus values are defined via CALCulate<Chn>:
LIMit:CONTrol:DOMain.
Tip: This command is complemented by CALCulate<Chn>:LIMit:RDOMain:
FORMat and CALCulate<Chn>:LIMit:RDOMain:SPACing.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
R&S ZVR/ZVABT Compatible Commands