721User Manual 1173.9163.02 ─ 55
In this case enable error logging for external devices (see "Log Errors"
on page 719), and search the Error Log Tab for self test error codes of the
device.
The device driver does not match exactly the device type. Probably the device
cannot be controlled properly. Measurements results are questionable.
Remote command:
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:DEFine
SYSTem:COMMunicate:RDEVice:PMETer:DELete
SYSTem:COMMunicate:RDEVice:PMETer:CATalog?
SYSTem:COMMunicate:RDEVice:PMETer:COUNt?
Scan Instruments
Performs an autodetection of "Known Devices" on the supported interfaces.
Note: Unintentional switchover to remote control
When using the NI-VISA library, ensure that the network analyzer itself is not listed as
a network device in the Measurement & Automation Explorer. Otherwise, "Scan Instru-
ments" initiates an identification query (*IDN?), causing the analyzer to close the
"External Power Meters" dialog (without executing "Scan Instruments") and to activate
the remote screen.
Remote command:
n/a
LAN Detection
Activate "LAN Detection" to include the LAN interface in the autodetection sequence
(see Scan Instruments).
Note: "LAN detection" only works for external devices on the IP subnet with the
R&S ZNB/ZNBT.
As a prerequisite, the R&S ZNB/ZNBT must accept incoming connections on UDP port
2473. If necessary, add a corresponding inbound rule to your firewall settings.
Remote command:
n/a
Add Device
Adds a new instrument to the list of "Configured Devices".
Setup Softtool