EasyManua.ls Logo

R&S ZNB - Page 931

R&S ZNB
1496 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
931User Manual 1173.9163.02 ─ 55
Usage: Event
Manual operation: See "DC Value" on page 320
CALCulate<Chn>:TRANsform:TIME:LPFRequency
Calculates the harmonic grid for low pass time domain transforms, keeping the stop
frequency and the number of points.
Tip: Use CALCulate<Chn>:TRANsform:TIME:LPASs if you wish to use one of the
other algorithms for calculating the grid.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example: See CALCulate<Chn>:TRANsform:TIME[:TYPE]
Usage: Event
Manual operation: See "DC Value" on page 320
CALCulate<Chn>:TRANsform:TIME:RESolution:EFACtor <REfactor>
Defines the resolution enhancement factor for the time domain transform.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<REfactor> Resolution enhancement factor.
Range: 1 to 10.
Increment: 0.1
*RST: 1 (no resolution enhancement)
Example: See CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam
Manual operation: See "Resolution Enh." on page 318
CALCulate<Chn>:TRANsform:TIME:SPAN <Span>
Defines the time span of the diagram in time domain.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Span> Time span of the diagram in time domain.
Range: 2E-012 s to 200 s.
Increment: 0.1 ns
*RST: 5E-009 s
Default unit: s
Example: See CALCulate<Chn>:TRANsform:TIME:CENTer
SCPI Command Reference