Command Reference
R&S
®
ZNB/ZNBT
962User Manual 1173.9163.02 ─ 55
Parameters for setting and query:
<CircuitModel> SC | SG
Possible circuit models (character data); see Table 8-8.
Example:
*RST; :CALC:TRAN:VNET:GLO:DEEM:PAR:C? SC
Query the default capacitance for ground loop deembedding.
The response is 1E-012 (1 pF).
CALC:TRAN:VNET:GLO:DEEM:PAR:C SC, 2.2E-12
Increase the capacitance to 2.2 pF.
Manual operation: See "Network" on page 608
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:G <CircuitModel>, <Conductance>
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:G? <CircuitModel>
Specifies the conductance value G in the different circuit models for ground loop
embedding.
Suffix:
<Ch>
.
Channel number.
<group> Port group (DUT) number.
If multiple port groups are configured (see SOURce<Ch>:
GROup<Grp>:PPORts) and CALCulate<Ch>:TRANsform:
VNETworks:GLOop:GROup is ON, then each port group can
have its own de-/embedding models.
Parameters:
<Conductance> Conductance G for the specified circuit model.
Range: -1kS to 1 kS.
Increment: 1 nS (1E-9 S)
*RST: 0 S
Default unit: Siemens (SI unit symbol: S)
Parameters for setting and query:
<CircuitModel> SG
Possible circuit models (character data); see Table 8-8.
Manual operation: See "Network" on page 608
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:L <CircuitModel>, <Inductance>
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:L? <CircuitModel>
Specifies the inductance value in the different circuit models for ground loop deembed-
ding.
Suffix:
<Ch>
.
Channel number.
SCPI Command Reference