Chapter 5
Maintenance and Repair
Trimble S, VX, SPS & RTS Service Manual5 - 22 P/N 57150002, Revision 5.0
F
4 Mount USB board Mount USB board in the
TCU attachment and
tighten the two screws
with a torque wrench
25 Ncm.
5 Connect cable Connect the cable from the
FPBTR board to the USB
board
6 Mount TCU
attachment with USB
Mount the TCU
attachment to the
instrument, see “TCU
attachment replacement”
on page 5-20.
7 Test USB function on
TCU attachment
See “Test USB function on
TCU attachment” on
page 6-14.
Item Performance Action Results Notes