Chapter 5
Maintenance and Repair
P/N 57150002, Revision 5.0 5 - 95 Trimble S, VX, SPS & RTS Service Manual
F
Replacement of Transceiver board
Performance
10 Check filter motor. See page 6-113.
11 Mount distance and
tracker unit cover.
See page 5-3.
12 Mount handle. Use a torx T30 screw
driver.
13 Perform offset
determination.
See page 6-82.
Item Performance Action Result Notes
1 Remove handle. Use a torx T30 screw
driver.
2 Remove distance unit
cover.
See page 5-3.
3 Disconnect
transceiver board
cables and screws.
Disconnect cables from:
A. SFC cable
B. Transmitter unit
C. Filter motor
D. Switch motor
E. Greywedge motor
F. Loosen receiver diode
clamp screw.
G. Remove screws.
1.3 mm allen key.
Item Performance Action Result Notes