Chapter 6
Adjustment and verification
P/N 57150002, Revision 5.0 6 - 113 Trimble S, VX, SPS & RTS Service Manual
F
Motor check
Fig. 6-128 Adjustment reference fibre
7 Exit software. A. Remove Scattered
light device with
adapter ring.
B. Click
Exit.
8 Perform offset
determination.
See page 6-120.
Item Performance Description Result Notes
1 Prepare instrument. A. Mount the instrument
on the Collimator
bench or on a stable
base.
B. Level the instrument.
2 Start up instrument /
PASS.
See page 7-3 how to set-up
the instrument.
3 Start service software. Select
Check motors from
EDM High Precision in
Maintenance & Service.
Item Performance Description Result Notes
A&C
B