Chapter 5
Maintenance and Repair
P/N 57150002, Revision 5.0 5 - 107 Trimble S, VX, SPS & RTS Service Manual
F
Distance unit - DR
• Laser Safety
• Differences to High Precision / DR Standard
• Replacement of Transmitter unit
• Replacement of Transceiver board
• Replacement of Switch motor
• Replacement of Greywedge motor
• Replacement of EDM Frame unit
Laser Safety
The Trimble S3 Total Station with distance unit DR is laser class 3R product See page 1-2
C
Warning – Avoid aiming the Class 3R Laser Pointer towards any person.
Service recommendation – Position the collimator equipment to avoid accidental aiming to any
person
Differences to High Precision / DR Standard
Differences compared to High Precision / DR Standard EDM. High Precision / DR Standard values
in brackets.
• Transmitter unit laser class 3R (class 2)
• Receiver fibre 100m (200m)
• Coax cable 250mm (65mm)
Note – 250 mm Coax cable must be used in DR EDM due to electrical disturbance
Note – Make sure that the coax cable goes free from the greywedge.