Chapter 6
Adjustment and verification
P/N 57150002, Revision 5.0 6 - 115 Trimble S, VX, SPS & RTS Service Manual
F
Verifications
check of signal/noise ratio, crosstalk, reference signal and distance measurement.
5 Check Switch motor. A. Select
Check path
switching
check box.
B. Click
Path switching
button.
C. Confirm the test results
by pressing
Yes or No
button.
An advice box
appears.
The test result
Test passed or
Test failed
will
be shown.
Observe the
movement of the
motor and listen
carefully at least 15-
20 times. You have to
hear a clear stop
noise of the filter
motor.
The switching
element moves until
the decision is made
6 Check greywedge
function.
A. Select
Check
greywedge
check box.
B. Click
Greywedge
button.
Grey wedge test
successful is
displayed when
the test is
successful.
Item Performance Description Result Notes
1 Prepare instrument. Mount the instrument on
the Collimator bench or
on a stable base.
2 Start up instrument /
PASS.
See page 7-3 how to set-
up the instrument.
3 Start service software. Select
Check Reference
fibre/signals
from EDM
High Precision
in
Maintenance & Service.
Item Performance Description Result Notes