Chapter 6
Adjustment and verification
P/N 57150002, Revision 5.0 6 - 19 Trimble S, VX, SPS & RTS Service Manual
F
Calibrate touch screen
Performance
Test
Performance
3 Install/Update
Application software
Item Performance Description Result Notes
1 Start up instrument
2 Calibrate touch screen A. Tap the start menu and
select Settings/control
panel/Stylus
B. Tap recalibrate in the
calibration tab
C. Follow the prompts on
the screen
Item Performance Description Result Notes
1 Start up instrument /
PASS.
See page 7-3 how to set-up
the instrument.
2 Test Control panel. A. Select
CU test from
Fixed CU in
Maintenance & Service.
B. Follow instructions.
Item Performance Action Result Notes