Chapter 6
Adjustment and verification
P/N 57150002, Revision 5.0 6 - 5 Trimble S, VX, SPS & RTS Service Manual
F
Servo side cover
Knob calibration
Trigger key function
Tools and equipment
Knob calibration
Performance
Trigger key function
Performance
Description Part number
PASS service software
Item Performance Description Result Notes
1 Start up instrument /
PASS.
See page 7-3 how to set-
up the instrument.
2 Calibrate knobs. A. Select
Knob calibration
from Knobs in
Maintenance & Service.
B. Follow instructions.
See page 3-14 for
information about
servo knob
calibration.
Item Performance Description Result Notes
1 Start up instrument /
PASS.
See page 7-3 how to set-up
the instrument.
2 Check trigger key
function.
A. Select
Trigger key from
Trigger key in
Maintenance & Service.
B. Follow instructions.