Chapter 6
Adjustment and verification
Trimble S, VX, SPS & RTS Service Manual6 - 166 P/N 57150002, Revision 5.0
F
Calibration of Video Camera
Performance
Item Performance Action Result Notes
1 Start up instrument /
PASS.
See Heading and Page
on Trimble S6 manual.
2 Perform Tilt sensor
offset.
See page 6-29.
3 Perform HA/VA
collimation
See page 6-47.
4 Perform Trunnion axis
correction
See page 6-48.
5 Run EDM offset
determination.
See page 6-82 and
page 6-120.
The distance is
important for the
camera module to
detect the ellipse.
6 Prepare instrument. A. Mount the
instrument on a
stable base.
(Collimator bench
or tripod).
B. Level the
instrument.
7 Start Service software. A. Select
Camera
calibration
from
Video in
Maintenance &
service
.
B. Connect the
instrument to the
PC and turn it on.
An image can be seen
in the dialogue box.