Chapter 6
Adjustment and verification
P/N 57150002, Revision 5.0 6 - 167 Trimble S, VX, SPS & RTS Service Manual
F
8
Setup of Calibration
target 1 & 2.
A. Select check box
Target set-up.
B. Click Laser pointer
ON the and aim
coarsely with the
red beam to the
centre of
Calibration target 1.
C. Click
Measure Face
I
button.
D. The instrument
turns to face II and
repeat aiming to
target 1. Click
Measure Face II
button.
E. Aim coarsely with
the Laser pointer to
the centre of
Calibration target 2.
F. Click
Measure Face
I
button.
G. The instrument
turns to face II and
repeat aiming to
target 2. Press the
Measure Face II
button.
H. If the instrument
could measure both
targets the message
Target setup
OK is
shown in results/
status.
On
Aiming target
1
the Measure
Face I
button is
active.
The measured
distance 1 is
displayed in the
status field. The
Measure Face II
button is active.
On
Aiming target
2
the Measure
Face I
button is
active.
The
Measure Face
II
button is active.
The measured
distance 2 is
displayed in the
status field.
It doesn’t matter in
which face the set-up
will take place.
During the two
measurements the
particular brightness
of the targets will be
measured and stored
for calibration. Then
the light condition has
to be stable.
The check box
Camera calibration is
active. The measured
calibration distances
are displayed in the
status field.
Item Performance Action Result Notes