Chapter 5
Maintenance and Repair
Trimble S, VX, SPS & RTS Service Manual5 - 148 P/N 57150002, Revision 5.0
F
What to do after replacement
This table is a short form instruction what to adjustments and checks/tests after repair/service. When
replacing a unit such as Tilt sensor more boards / modules might need to be replaced.
Board / module replaced Adjustment Check / test
All instruments
After service / repair / calibration Run all tests in Final Test &
Report
Base instrument
Distance / Tracker unit cover PASS, HA/VA Collimation
test
Accelerometers PASS, Servo & Angle system
calibration, cleanliness test
Base unit PASS, Servo & Angle
calibration, cleanliness test &
Foot connectors test
BRC board PASS, Internal battery, Radio
range test (robotic)
Face 2 panel Optical plummet PASS, Face 2 panel
FFB board (Hirose connectors) PASS, Foot connectors test
Horizontal angle disc PASS, Servo & Angle
calibration, cleanliness test
Horizontal SRC board incl.
magnets
PASS, Servo & Angle
calibration, cleanliness test
Horizontal MSB board incl. motor
winding
PASS, Hardware Angle, MSB
board replacement, Servo &
Angle calibration, cleanliness
test
Regreasing of vertical axes PASS, Tilt sensor offset, HA/
VA collimation correction,
Trunnion axis correction,
Vertical oscillation
compensation (The Vertical
Friction Compensation should
be enabled on all instruments
and the vertical friction
constant should be below
75000 after regreasing, Tracker
collimation, Radio and camera
calibration on VX instruments.