Chapter 6
Adjustment and verification
Trimble S, VX, SPS & RTS Service Manual6 - 18 P/N 57150002, Revision 5.0
F
Set Serial number
Performance
Install Application software
Performance
Item Action Performance Results Notes
1 Start instrument In the control panel go to
Start /
Programs / Utilities /
Command Promt.
Note – If the wrong
serial number is
entered the Quimby
board must be
replaced!
2 Enter serial number Type
awcutil -
setserial:XXXXXXXX
Press return
The serial
number is set.
XXXXXXXX =
Serial Number
Item Performance Action Result Notes
1 Set up instrument A. Connect instrument
to power supply
B. Connect control
panel USB to PC
USB
Fig. 6-13 S3 Set up
2 Download Application
software
A. Go to http://
www.trimble.com
B. Select product
C. Select support
D. Select Downloads
PC USB
+12V