Chapter 6
Adjustment and verification
P/N 57150002, Revision 5.0 6 - 129 Trimble S, VX, SPS & RTS Service Manual
F
4 Change Face. Click Switch face Instrument will
change to the
opposite face.
5 Aim to the collimator
in first face.
A. Aim to the collimator.
B. Click
Measure Face1.
The angle values
for Hz I and V I
are displayed
together with
CHz and CV i.
The values for CHz
and CV are
representing the
values for “c” and “i”,
“c” means the
horizontal
collimation error and
“i” means the vertical
index error.
6 Decision tolerance
passed.
If tolerance passed
(“Yes”) go on with tracker
receiver, see page 6-130.
If tolerance failed (“No”)
go back to reticle
adjustment, see page 6-33.
Chz (c) 93cc
(30”)
CV (i) 93cc
(30”)
Item Performance Description Result Notes