Chapter 6
Adjustment and verification
Trimble S, VX, SPS & RTS Service Manual6 - 132 P/N 57150002, Revision 5.0
F
6 Decision: Tolerance
passed.
If tolerance passed
(“Yes”) go on with final
calibration.
If tolerance failed (“No”)
go on with tracker
receiver adjustment, see
page 6-133.
See Fig. 6-142
on page 6-131 :
Chz 93cc
(30”)
CV 93cc
(30”)
c- cT 123cc
(40”)
i - iT 123cc
(40”)
An alignment error of
40” corresponds to 20
mm on 100 meter.