Messages
Message
Location
ExplanationWhattodo
●Alarmvolume
changed
●al.areaThenetworkconnectionislost,
andthelocalalarmvolumeis
increased.
●Readjustvolumeifdesired.
●Alarmsacknowl-
edgedremotely
●al.areaThealarmshavebeen
acknowledgedfromthecentral
station.
●Noactionrequiredbutyou
maywishtoconrminalarm
historywhichalarmhasbeen
acknowledged.
●Alarmsaudiopaused
remotely
●al.areaThealarmswereremotely
pausedfromthecentralstation.
●Youcanactivatethealarms
bypressingtheaudiopause
key.
●Alarmsaudiopaused
fromtelemetry
●al.areaThealarmswereremotely
pausedfromthetelemetry
transmitter.
●Youcanactivatethealarms
bypressingtheaudiopause
key.
●Allmonitors
disconnected
●al.areaThemonitorisdisconnected
fromthenetwork.
●Re-establishconnection.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Application
error:iPanel
●al.areaConnectiontoiPanelwaslost.●Re-establishtheconnection.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Applicationerror:pdf●al.areaPDFviewerclosesunexpectedly.●TrytoopentheMUSE12SL
reportagain.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Application
error:Webmin
●al.areaWebminapplicationcloses
unexpectedly.
●TrytoopenWebminagain.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●AudiosetOFF
remotely
●al.areaThealarmshavebeenturnedoff
fromthecentralstation.
●Youcanchangetheselection
throughAlarmSetup>
Audible&Visual.
●Barcodescanned●al.areaAlldatahasbeensuccessfully
storedtothemonitor.
●Noactionrequired.
●Barcodetoolong●al.areaThemaximumlengthofthe
barcodehasbeenexceeded.
●Verifytheinformationreadby
thebarcodereaderandeditif
necessary.
●BatteryAfailure
●BatteryBfailure
●al.areaBatteryAorBisfaulty.●Replacethebatteryif
necessary.
●Batteryfailure●al.areaThemonitorbatteryisfaulty.●Replacethebatteryif
necessary.
506CARESCAPEModularMonitors2062971-001