Messages
Message
Location
ExplanationWhattodo
●BatteryA
temperaturehigh
●BatteryB
temperaturehigh
●al.area
BatteryAorBtemperatureistoo
high.
●Replacethebattery.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Batterytemperature
high
●al.areaThebattery’stemperatureistoo
high.
●Replacethebattery.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●<Bed>Monitor
disconnected
●al.areaThemonitorwithalarm
noticationenabledis
disconnectedfromthenetwork.
●Re-establishtheconnection.
●CallService:Text(s)
missing
●al.areaThesoftwaretextismissingin
thislanguage;thetextlemay
becorrupted.
●Contactauthorizedservice
personnel.
●Caseended●al.areaORandPACUsoftware
packages:Thecurrentcase
hasjustbeenended.
●Noactionrequired.
●Casestarted●al.areaORandPACUsoftware
packages:Anewcasehas
justbeenstarted.
●Noactionrequired.
●CheckPDMbattery●al.areaPDMbatteryisnotworking
properly.
●Checkandreplaceorremove
thebattery.
●ConditionbatteryA
●ConditionbatteryB
●al.areaBatteryAorBisnotworking
properly.
●Replaceorremovethe
battery.
●Conditionmonitor
battery
●al.areaBatteryisnotworkingproperly.●Replaceorremovethe
battery.
●Conguration
change(s)
●al.areaTheloadedcongurationhas
changedfromthepreviousone.
●Noactionrequired,but
youmaywishtocheckthe
settings.
●Conguration
changes.Restart
required.
●al.areaThecongurationhaschanged.●Restartthemonitor.
●Congurationerror(s)●al.areaOneormoreerrorshavebeen
detectedintheconguration.
●Contactauthorizedservice
personnel.
●Connecting
Measurement
●al.areaAnacquisitionmodulehasbeen
connected.
●Noactionrequired.
●Connectingtelemetry
transmitter
●al.areaThemonitorisconnectingtoa
telemetrytransmitter.
●Noactionrequired.
●Countdowntimer
expired
●al.areaUsersetcountdowntimerhas
expired.
●Resetthetimerifrequired.
●Enteringstandby●al.areaActivatestandbyhasbeen
selected.
●Noactionrequired.
2062971-001CARESCAPEModularMonitors507