Loading...
Messages
Message
Location
ExplanationWhattodo
License(s)expiredal.areaOneormoretriallicenseshave
expired.
Contactauthorizedservice
personnel.
Loadingcompletedal.areaLoadingormerginga
case/patientfromtheS/5
Networkwascompleted.
Noactionrequired.
Loadingfailedal.area
Loadingacase/patientfroman
acquisitionmoduleornetwork
hasbeeninterrupted.
Checkdeviceornetwork
cableconnections.
Loadingfailed.
Wrongpatient
al.area
MergingacasefromtheS/5
Networkwasrefusedbythe
centralstation.
CheckthattheLastName,
FirstNameandMedical
RecordNumberareidentical
onthemonitorandcentral
station,orthatthemonitor
hasnoinformationinthese
elds.
Loadingfrom
network
al.areaPatientdataisbeingloaded
fromthenetwork.
Noactionrequired.
LoadingfromPDM/
LoadingfromTram
al.areaPatientdataisbeingloaded
fromanacquisitionmodule.
Wait.
Modulevoltagelowal.areaParametersmaynotbeworking
properlyduetoatechnicalfault
inthemonitor.
Checkthepatientstatus.
Contactauthorizedservice
personnel.
Monitorbattery
empty!
al.areaThemonitorisusedonbattery
power,andthereislessthan5
minutesofchargeleft.
Chargethebatterybyusing
themonitoronmainspower.
Monitorbatterylowal.areaThemonitorisusedonbattery
power,andthereislessthan20
minutesofchargeleft.
Chargethebatterybyusing
themonitoronmainspower.
Monitordisconnectedal.areaThealarmingmonitoris
disconnectedfromthenetwork.
Re-establishtheconnection.
Monitorpowering
down!
al.areaThemonitorisusedonbattery
powerandthereislessthan10
secondsofchargeleft.
Chargethebatterybyusing
themonitoronmainspower.
Network:<central
name>
al.areaConnectiontoaniCentralis
established.
Noactionrequired.
Networkdownal.areaCARESCAPENetworkconnection
hasfailed.
B650,B450:Ifthemonitor
isusedwithWLANoption,it
isinshadowregionandnot
connectedtothenetwork.
Trytore-establishthe
connection.
Contactauthorizedservice
personnel.
Networkdown:
<centralname>
al.areaThemonitorhaslostconnection
toaniCentralintheS/5Network.
Checkthenetwork
connections.
Nobatterybackupin
monitor
al.areaThemonitorhasthebattery
optionbutthereisnobattery
inserted.
Insertabattery.
2062971-001CARESCAPEModularMonitors509