11.1.2.1.1 Distance 2 Points
1.
To measure the distance between two points, select the [Dist. 2Point] item in the menu
area. A cursor appears on the screen.
2. Use the trackball to move the cursor to the start point of the measurement and press the
right or left trackball key [Set] to fix the marker. A second cursor appears.
3. Move the second cursor to the second point of the measurement and press [Set] again.
Note
To re-adjust the start point, press the upper trackball key [Change] before completing the
measurement. This alternates the control from one cursor to the other.
11.1.2.1.2 Distance 2 Lines
1. To measure the distance between two lines, select the [Dist. 2Line] item in the menu
area. A line appears on the screen.
2. Move the line to the start point of the measurement and press the right or left trackball
key [Set].
Note
To re-adjust the start point, press the upper trackball key [Change].
3. Move the trackball to adjust the angle and then press [Set] again. A second line (parallel
to the first one) appears.
4. Move this line using the trackball to the end point of the measurement and then press
[Set].
11.1.2.1.3 Length Trace
1. To measure the distance between two points using trace, select the [Length Trace] item
in the menu area. A cursor appears on the screen.
2. Move the cursor to the start of the measurement and press the right or left trackball key
[Set] to fix the marker. Begin trace.
Note
To re-adjust the traced line, press the upper trackball key [Undo] repeatedly.
3. Trace to the end and press the [Set] key again to fix the mark.
11.1.2.1.4 Length Point
1. To measure the distance between several points (as much as desired), select the
[Length Point] item in the menu area. A cursor appears on the screen.
2. Move the cursor to the start of the measurement and press the right or left trackball key
[Set] to fix the marker.
3. Move the trackball again to adjust the next line between two points and then press [Set]
again.
Note
To re-adjust a line, press the upper trackball key [Undo] repeatedly.
4. Set as much points as required in the same manner.
5. To finish the measurement and to display the result, press the [Set] key once again.
11.1.2.1.5 Stenosis %Distance
1. To measure a Stenosis, select the [Stenosis %Dist] item in the menu area. A cursor
appears on the screen.
2. Perform the measurement of the outer distance of the stenosis using the trackball and
the right or left trackball key [Set]. A second cursor appears.
3. Perform the measurement of the inner distance of the stenosis and press the [Set] key.
Measurements and Patient Worksheets (Reports)