1. Press the [Calc] key on the control panel.
2.
Select the item [PAP].
3. Select the appropriate measurement parameter. For example: select [VPD].
4. If necessary, select the [Angle] and the [Baseline].
5. Perform the measurement by using the trackball and the right or left trackball key [Set].
11.2.5.34 HR (Heart Rate)
The method of the measurement is the same as that of measuring the Heart Rate in M mode.
operation review:
HR (Heart Rate)
(chapter
'HR (Heart Rate)'
on page 11-73
)
11.2.5.35 Cardiology - Worksheet
Press the [Report] hardkey on the control panel to view the report that contains detailed results
of cardio calculations.
Measurements and Patient Worksheets (Reports)
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4 11-81