Coded PI
Coded PI: CIS
Coded PI: CCIS
Acc. power range: 1 - 100%
Scan angle: taken from 2D
GAIN range: +15 to -15 dB
Gray scale values: 32 bit
SRI taken from 2D
Persistance filter: 8 steps (pre)
S./PRI 1.00, 1.50, 2.00, 3.00, 4.00..................15.00
Line Density 3 steps (pre) low, norm, high
Enhance: 6 steps (pre) 0 - 5
Gray maps: 18
Tint maps: 15
Dynamic: 12 different dynamic curves C1 -C12
Accumulation: Off, 0.20, 0.40, 0.60, 0.80, 1.60, 3.20 Infinite
Background: 0, 1, 2
Time delay: 0, 0.5, 1, 2, 3, ..........10
16.10.11 Elastography
Screen formats: 2D+Elastography (Single, Dual, Quad)
Display modes: Simultaneous dual mode: 2D/2D+ Elastography
Acc. power range: 1 - 100%
Scan angle: taken from 2D
Soft Compress: Range: 0 - 9; step size 1
Hard Compress: Range: 0 - 9; step size 1
Frequency: steps: penet/norm/resol
Transparency: Range: 0 - 255; 0: not transp.; 255: full transp.; step size 5
Persistance: Range: 1-9; step size 1
Line Dens: Range: 1 - 2
Filter Axial: Range: 1-9; step size 1
Filter Lateral: Range: 1-21; step size 2
Window Length: Range: 8-25; step size 1
Window Step: Range: 1-max; step size 1(max = 0.8 * current WindowLength)
Frame Reject: Range: 0-255; step size 5
Pixel Reject: Range: 0-255; step size 5
AGC Filter axial: Range: 1-63; step size 2
Technical Data / Information
16-24
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4