1. Press the [Patient] key on the control panel, select the [CARD] page and enter all patient
information for Cardio calculations. For details refer to: Entering Patient Data
(chapter
'Entering Patient Data'
on page 4-10
)
Note
To cancel all calculations performed before and to start new measurement, press this key and
select [End Exam] or [Clear Exam].
2. Select [Start Exam].
2. Make sure that the probe and application are selected properly. If another application is
selected, press the [Probe]
key on the control panel, and change it to Cardiology. For details
refer to:
Probe/Program Selection
(chapter
'Probe/Program Selection'
on page 4-7
)
11.2.5.3 Cardiology Calculations in 2D Mode
LV Simpson
(chapter
'LV Simpson'
on page 11-62
)
Vol A/L (Volume Area/Length)
(chapter
'Vol A/L (Volume Area/Length)'
on page 11-62
)
LV (Left Ventricle)
(chapter
'LV (Left Ventricle)'
on page 11-64
)
LV Mass
(chapter
'LV Mass'
on page 11-64
)
LVOT- or RVOT Diameter
(chapter
'LVOT- or RVOT Diameter'
on page 11-65
)
MV (Mitral Valve)
(charter
'MV (Mitral Valve)'
on page 11-67
)
TV (Tricuspid Valve)
(chapter
'TV (Tricuspid Valve)'
on page 11-67
)
AV/LA (Aortic Valve/Left Atrium)
(charter
'AV/LA (Aortic Valve/Left Atrium)'
on page 11-
68
)
PV (Pulmonary Valve)
(charter
'PV (Pulmonary Valve)'
on page 11-68
)
Measurements and Patient Worksheets (Reports)
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4 11-61