Recommended separation distances between portable and mobile RF communications equipment and the
Voluson© S6/S8
0.1 11.07 0.37 0.74
1 35.00 1.17 2.33
10 110.68 3.70 7.40
100 350.00 11.70 23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using equation applicable to the frequency of transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1 At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.