Command Reference
R&S
®
ZNB/ZNBT
1288User Manual 1173.9163.02 ─ 55
SOURce<Ch>:POWer<PhyPt>:STATe............................................................................1296
SOURce<Ch>:POWer<PhyPt>:STARt........................................................................... 1296
SOURce<Ch>:POWer<PhyPt>:STOP............................................................................1296
SOURce:POWer:GENerator:SDELay <SettlingDelay>
If external generators are used in the measurement, this defines the minimum settling
delay before the external generators are considered to be ready, i.e. the time between
setting the generator to the requested frequency and power and continuing with the
measurement.
This is a global setting.
Parameters:
<SettlingDelay> Settling delay per sweep point.
Factory default is 10 µs (= 10
-5
s).
Default unit: s
Manual operation: See "External Generators > Minimum Settling Delay"
on page 709
SOURce:POWer:REDuce:SDELay <SettlingDelay>
Deprecated; superseded by SOURce:POWer:SWEepend:SDELay
If power reduction at sweep end is enabled (SOURce<Ch>:POWer<PhyPt>:
REDuce[:STATe] ON) this defines the settling delay at subsequent sweep starts, i.e.
the time between power-up and sweep start when a new sweep is requested.
This is a global setting.
Parameters:
<SettlingDelay> Settling delay
Default unit: s
Example: See SOURce:POWer:SWEepend:MODE on page 1288
SOURce:POWer:SWEepend:MODE <SweepEndModes>
Selects the power mode at sweep end.
This is a global setting.
Parameters:
<SweepEndModes> AUTO | REDuce | KEEP
AUTO: at sweep end restore the power at sweep start
REDuce: at sweep end reduce the output power as if the chan-
nel base power was set to its minimum possible value
KEEP: at sweep end keep the power at its current level
SCPI Command Reference