GUI Reference
R&S
®
ZNB/ZNBT
588User Manual 1173.9163.02 ─ 55
The "2-Port Data" network is defined by means of an s2p file (see Select File...).
For deembedding, the s2p file can also be generated by a third-party fixture model-
ing tool (see "Fixture Tool" on page 589).
The remaining networks are defined by lumped elements. Their parameters are
displayed below the graphical list.
Tip: Drag and drop the network symbols in horizontal or vertical direction to switch to
the next symbol. The lumped elements are numbered from top to bottom.
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>[:
STATe]
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
TNDefinition
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:C<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:G<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:L<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:R<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>[:
STATe]
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
TNDefinition
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
PARameters:C<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
PARameters:G<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
PARameters:L<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
PARameters:R<Cmp>
Select File...
"Select File..." is enabled as long as the "2-Port Data" network is selected. This net-
work is defined by its S-parameters stored in a two-port Touchstone file (*.s2p). No
additional parameters are required.
In case the port number conventions of the loaded two-port Touchstone file differ from
network analyzer conventions (port 1 on the left, i.e. on the analyzer side; port 2 on the
right, i.e. on the DUT side), it is possible to "Swap Gates". The analyzer will inter-
change the port numbers (e.g. S
12
--> S
21
) when loading the file.
Note: The loaded file is stored in the active recall set. Recall sets contain the full
(de-)embedding data so that they can be transferred to other instruments.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:SENDed:DEEMbedding<PhyPt>
MMEMory:LOAD:VNETworks<Ch>:SENDed:EMBedding<PhyPt>
Offset Embed Softtool