Command Reference
R&S
®
ZNB/ZNBT
828User Manual 1173.9163.02 ─ 55
Parameters:
<Boolean> 1 (ON, true) if the test fixture is connected to port <Ph_pt>, 0
(OFF, false) otherwise
Example: See Chapter 9.2.8, "Fixture Modeling", on page 1415
Manual operation: See "Active" on page 594
CALCulate:FMODel:ISD<Ph_pt>:FTIMe:DUT <FltDutAndLeadIns>
If CALCulate:FMODel:ISD<Ph_pt>:FTIMe:OVERride is set to TRUE, this com-
mand allows to set the flight time for DUT + Lead-ins manually.
This is equivalent to setting the leadin_dut_time batch mode parameter of the ISD
tool to a numeric value (not auto).
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<FltDutAndLeadIns> Default unit: s
Manual operation: See "Flt Tm for DUT + Lead Ins" on page 598
CALCulate:FMODel:ISD<Ph_pt>:FTIMe:OVERride <Boolean>
If set to true, the flight time for DUT + lead-in is specified using CALCulate:FMODel:
ISD<Ph_pt>:FTIMe:DUT. Otherwise, it is calculated automatically.
True is equivalent to setting the leadin_dut_time batch mode parameter of the ISD
tool to auto.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<Boolean>
Manual operation: See "Automatic Flt Tm for DUT + Lead Ins" on page 598
CALCulate:FMODel:ISD<Ph_pt>:OPERation <Operation>
Defines the execution mode of the ISD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<Operation> FAST | ACCurate
ACCurate: normal execution mode
FAST: reduces the execution time to ~50%
In many cases the FAST mode can be enabled with only little
loss of accuracy.
SCPI Command Reference