Command Reference
R&S
®
ZNB/ZNBT
827User Manual 1173.9163.02 ─ 55
Can only be set to ON, if a recent version of the ISD tool (from 2019-12 or later) is
used. ON correspons to add_dc=1 in the tool's batch mode.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<Boolean> OFF (0)
No DC data added, i.e. the touchstone data reach down to the
start frequency of the instrument sweep.
ON (1)
The ISD tool adds (extrapolated) S-parameter data for f = 0.
*RST: OFF (0)
Manual operation: See "DC Extrapolation" on page 600
CALCulate:FMODel:ISD<Ph_pt>:DUT:MEASure
For a fixture modeling with the ISD tool, this command allows to measure DUT + Fix-
ture at the active ports (see CALCulate:FMODel:ISD<Ph_pt>:DUT[:STATe])
Suffix:
<Ph_pt>
.
This suffix is ignored.
Example: See Chapter 9.2.8, "Fixture Modeling", on page 1415
Usage: Event
Manual operation: See "Measure" on page 594
CALCulate:FMODel:ISD<Ph_pt>:DUT:TYPE <DUTType>
Defines whether the DUT is active or passive.
Corresponds to the active_dut batch mode parameter of the ISD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<DUTType> PASSive | ACTive
Manual operation: See "DUT Type" on page 598
CALCulate:FMODel:ISD<Ph_pt>:DUT[:STATe] <Boolean>
For a fixture modeling with the ISD tool, this command allows to specify the ports to
which the test fixture is connected.
This has to be defined before measuring DUT + Fixture (using CALCulate:FMODel:
ISD<Ph_pt>:DUT:MEASure).
Suffix:
<Ph_pt>
.
Physical port number
SCPI Command Reference