Command Reference
R&S
®
ZNB/ZNBT
840User Manual 1173.9163.02 ─ 55
Usage: Event
Options: R&S ZNB/ZNBT-K231
Manual operation: See "Clear All 1L Measurements/Clear Selected 1L Measure-
ments" on page 617
CALCulate:FMODel:DELT<Ph_pt>:M1L:CACHe:CLEar:SELected <String>
If the 1-length method is used (CALCulate:FMODel:DELT<Ph_pt>:METHod M1L),
the VNA firmware builds up a cache of coupon measurements (Touchstone files) in the
DeltaL\1LMeasurementCache subfolder of the user data folder
C:\Users\Public\Documents\Rohde-Schwarz\Vna.
This command deletes the selected file from this cache.
Suffix:
<Ph_pt>
.
Physical port number
Setting parameters:
<String> Path to the Touchstone file, relative to the cache directory.
Usage: Setting only
Options: R&S ZNB/ZNBT-K231
Manual operation: See "Clear All 1L Measurements/Clear Selected 1L Measure-
ments" on page 617
CALCulate:FMODel:DELT<Ph_pt>:M1L:FILename <String>
CALCulate:FMODel:DELT<Ph_pt>:M2L:FILename <String>
CALCulate:FMODel:DELT<Ph_pt>:M3L:FILename <String>
Loads coupon measurement results for the 1st/2nd/3rd line from a s2p or s4p Touch-
stone file.
Suffix:
<Ph_pt>
.
This suffix is ignored
Parameters:
<String> Path to the Touchstone file, either absolute or relative to the cur-
rent directory (see MMEMory:CDIRectory).
Options: R&S ZNB/ZNBT-K231
Manual operation: See "Load File" on page 616
CALCulate:FMODel:DELT<Ph_pt>:M1L:LENGth <Length>
CALCulate:FMODel:DELT<Ph_pt>:M2L:LENGth <Length>
CALCulate:FMODel:DELT<Ph_pt>:M3L:LENGth <Length>
Specifies the length of the 1st/2nd/3rd line in Delta-L measurements (CALCu
late:FMODel:DELT<Ph_pt>:M1L|M2L|M3L:MEASure).
SCPI Command Reference