Command Reference
R&S
®
ZNB/ZNBT
974User Manual 1173.9163.02 ─ 55
Suffix:
<Ch>
.
Channel number
<ListId> Index of the affected port set (see CALCulate<Ch>:
TRANsform:VNETworks:PPAir:EMBedding<ListId>:
DEFine)
Parameters:
<Boolean> OFF (0): Embedding inactive
ON (1): Embedding active
*RST: OFF (0)
Example:
*RST; CALC:TRAN:VNET:PPA:EMB:DEF 1,2,3,4
Define a port pair configuration with port pairs (1,2) and (3,4).
CALC:TRAN:VNET:PPA:EMB1:TND CSSL
Select the Serial Cs, shunt L circuit model for the first port pair.
CALC:TRAN:VNET:PPA:EMB1:PAR:
R3 CSSL, 2.2E+3; CALC:TRAN:VNET:PPA:EMB1 ON
Increase the resistance R3 for the Serial Cs, shunt L circuit
model to 2.2 kΩ and enable deembedding.
CALC:TRAN:VNET:PPA:EMB:DEL
Delete the port pair configuration.
Manual operation: See "Active" on page 573
CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<ListId>:DEFine
<PP_First>,<PP_Second>,<PP_First>,...
Creates one or more port pairs for port set embedding. The command can be used
repeatedly to extend or (partially) overwrite the list of port sets for embedding.
See CALCulate<Ch>:TRANsform:VNETworks:PSET:EMBedding<ListId>:
DEFine for general port set definition.
Suffix:
<Ch>
.
Channel number
<ListId> Index of the defined port set within the channel's overall list of
port sets for embedding.
Port sets for embedding must be numbered consecutively, i.e.
port set <ListId> can only be created if port set <ListId>-1
already exists. If several port pairs are specified, <ListId> is the
number of the first port pair to be created.
Parameters:
<PP_First>,
<PP_Second>,
<PP_First>, ...
Sequence of port pairs, each one consisting of two different
ports.
The port pairs don't have to be disjoint.
Example: See CALCulate<Ch>:TRANsform:VNETworks:PPAir:
EMBedding<ListId>[:STATe]
Usage: Setting only
SCPI Command Reference