Chapter 6
Adjustment and verification
Trimble S, VX, SPS & RTS Service Manual6 - 46 P/N 57150002, Revision 5.0
F
Fig. 6-36 Loosen reticle screws
Fig. 6-37 Horizontal reticle adjustment
7 Exit Reticle
adjustment software.
A. Click
Exit button.
B. Click
OK to confirm
Run reticle collimation
text box, see Fig. 6-39
on page 6-47 .
C. Click OK to confirm
Remote tools from
instrument, see Fig. 6-
38 on page 6-47 .
D. Run HA/VA
collimation, see
“Reticle collimation
(HA/VA collimation)”
on page 6-47.
Item Performance Description Result Notes