Chapter 6
Adjustment and verification
Trimble S, VX, SPS & RTS Service Manual6 - 100 P/N 57150002, Revision 5.0
F
Fig. 6-110 EDM base screw position and adjustment
11 Check transmitter
beam (laser pointer)
focus.
A.Select
Switch filter mo-
tor HIGH/LOW
check
box.
B. Click Filter motor
HIGH.
C. Aim to a 30 meter
target.
The transmitter
spot at a target
in 30 meter
distance has to
be 20 mm.
The evaluation of the
transmitter spot,see
Fig. 6-109 on page 6-
99 has to be done
under dark light
condition by close
observation to see the
complete
beamspread.
If the transmitter spot
is out of focus the
Transmitter unit
needs to be replaced.
Item Performance Description Result Notes
A&B
C
C
Vertical
Horizontal
B