Chapter 6
Adjustment and verification
P/N 57150002, Revision 5.0 6 - 169 Trimble S, VX, SPS & RTS Service Manual
F
9 Calibration of Video-
camera.
A. Select Camera
calibration
check
box.
B. Click
Start for
scanning the targets
automatically. No
aiming of targets is
needed, because the
position is stored
during the setup.
C. The calibration
result is shown in
the result/status
field.
D. The single
calibration values
are shown in the
field program
status/messages
with tolerances.
The
Start button
for
Scanning is
active.
After successful
completing, the
calibration result
OK will appear in
the status field.
If the calibration
fails the constants
which are out of
tolerance will be
shown in
program status.
See “Calibration
values out of
tolerance:” on
page 6-172.
The laser pointer can
be used for control or
switched off.
The status bar starts to
run for approximately
20 minutes scanning
the target in 2 faces in
sequence. If the
calibration is failing, a
message box will
appear and the
procedure has to be
repeated.
Use a well charged
battery for the
calibration process
and check the Power
Supply status in
advance!
Item Performance Action Result Notes