Adjustment and verification
Trimble S, VX, SPS & RTS Service Manual6 - 172 P/N 57150002, Revision 5.0
F
Error messages Calibration values out
of tolerance:
x0
out of tolerance
y0 out of tolerance
ck out of tolerance
kappa out of tolerance
s0 out of tolerance
c0 out of tolerance
z0 out of tolerance
k1 out of tolerance
m0 out of tolerance
What’s to do?
Check position of camera module to
optical axis and mechanical supports.
Repeat focusing of video front lens and
check the sharpness.
Check/replace video front lens and repeat
focusing.
Item Performance Action Result Notes