Chapter 6
Adjustment and verification
P/N 57150002, Revision 5.0 6 - 173 Trimble S, VX, SPS & RTS Service Manual
F
7 Test measurement of
Video calibration
A. Select Test
measurement
check
box.
B. Check the
calibration
constants with a
Test measurement
to target 1
pressing
the Start button.
C. Move the centre of
the black dot
between the two red
circles lines as
shown. Confirm the
message box and
click with the
cursor on the dot to
start the first
measurement, see
Fig. 6-179 on page
6-174 .
D. The instrument will
turn to face II,
confirm the
message box and
select the dot in
face II with the
cursor again, see
Fig. 6-180 on page
6-175 .
E. After successful
completing the test,
the collimation
errors will be
displayed in results,
see Fig. 6-181 on
page 6-176 .
The
Start button
for
Test
measurement to
to target
is active.
Aim target 1
The dot will be
detected by the
software, shown
with a green
circle around.
The calculation
of the collimation
errors in
horizontal (HA)
and vertical (VA)
direction will be
done.
The test measurement
can be done
separately from a
complete calibration
procedure to check
the camera status. As
pre-condition the
Target set up has to be
done complete to run
the measurement.
Check the position of
the dot in face II
insight the two red
circles!
Specification:
HA < 0,02500 gon
VA < 0,02500 gon
Item Performance Action Result Notes