EasyManuals Logo

Barco S3 User Manual

Barco S3
538 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
3. General
3.7 Installation requirements
Environment conditions
Table below summarizes the physical environment in which the E vent Master devices an d their controllers may be s afely operated
or stored.
Environmen t
Operating Non-Operating
Ambient Tempe rature
0°C (32°F) to 40°C (104°F) -10°C (14°F) to 60°C (140°F)
Air cleanliness
Clean ofce environment (equivalent with cleanroom
standard ISO 14644-1 ISO Class 9)
n.a.
Humidity
5% to 85% R H Non-condensed 0% to 95% RH N on-Condensed
Altitude -60 (-197Ft) to 3000m (9843Ft) -60 (-197Ft) to 10000m (32810Ft)
CAUTION: Let the Event Master devices and their con trollers acclimate to the am bient temperature after un-
packing. E n sure that the humid ity is within the specication. Neglecting this may result in startup failure o f
these d evices.
Software package
This version of the User’s Guide is based on software version 03.00.00.
Verify that the Event Master device is loaded with the latest software version available on the Barco web site
(URL:h
ttps://ww w.barco.com/). R efer to the c hapte r "Up dating rmware", page 255 for more information about
the device upgrading software.
Event Master Toolset Software (EM GUI)
The EM GUI software must be at least:
Version 03.00.00
34
R5905948 EVENT MASTER DEVICES 17/07/2017

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Barco S3 and is the answer not in the manual?

Barco S3 Specifications

General IconGeneral
BrandBarco
ModelS3
CategoryComputer Hardware
LanguageEnglish

Related product manuals