3 VMA Mode
3.7 Power Stat CCDF Measurement
–
MEASured
–
GAUSsian
–
REFerence
Remote
Command
:CALCulate:PSTatistic:MARKer[1]|2|…|12:TRACe MEASured | GAUSsian | REFerence
:CALCulate:PSTatistic:MARKer[1]|2|…|12:TRACe?
Example
:CALC:PST:MARK3:TRAC MEAS
:CALC:PST:MARK:TRACE?
Preset
MEASured
State Saved Yes
Range
MEASured|GAUSsian|REFerence
Marker Settings Diagram
Lets you configure the Marker system using a visual utility. This is the same as
"Marker Settings Diagram" on page 1138 in Settings.
3.7.8 Meas Setup
Contains functions for setting up the measurement parameters and also contains
functions for setting up parameters global to all measurements in the mode.
3.7.8.1 Settings
Contains frequently used functions to which you will want the fastest access.
Counts
Sets the accumulated number of sampling points for data acquisition. The range is
1.000 kpt (k point) to 2.00000 Gpt (G point) with 1 kpt resolution.
Remote Command
[:SENSe]:PSTatistic:COUNts <integer>
[:SENSe]:PSTatistic:COUNts?
Example
:PST:COUN 5001
:PST:COUN?
Couplings Coupled to "Meas Cycles" on page 1141, by:
Counts = Meas Cycles * SamplingFrequency * "Meas Interval" on page 1141
Preset 10000000
Vector Modulation Analyzer Mode User's &Programmer's Reference 1140