9 Programming the Instrument
9.3 SCPI Operation and Results Query
9.3 SCPI Operation and Results Query
You can use SCPI commands for remote control of measurements and querying of
measurement results data. There are several alternative commands you can use to
control the measurement, depending on how you wish to operate the instrument.
There are also a number of queries that you can use to extract the measurement
data.
In this section, “Mode” refers to a Measurement Application, for example, Spectrum
Analyzer or 5G NR.
9.3.1 Mode Control
Use either :INSTrument:SELect or :INSTrument:NSELect to select the Mode.
See "Mode" on page 91.
:INSTrument:CONFigure causes a Mode and Measurement switch at the same
time. This results in faster overall switching than sending :INSTrument:SELect
and :CONFigure separately, as described in "Mode" on page 91.
9.3.2 Measurement Control
This section describes the measurement control commands listed below, and their
functions.
"CONFigure" on
page 2551
Switches to the desired measurement. Presets all measurement settings to their
defaults, unless:NDEFault is specified
"INITiate" on
page 2552
Starts the measurement
"FETCh" on page
2552
Queries the data without starting the measurement. If a measurement is already
in progress, waits for completion
"READ" on page
2553
Starts the measurement with the current settings and queries the data
"MEASure" on
page 2554
Switches to the desired measurement, presets all measurement settings to their
defaults, starts the measurement and queries the data
The relationship between the command forms is illustrated in the diagram below.
:FETCh, :READ and :MEASure are queries that return measurement data.
Vector Modulation Analyzer Mode User's &Programmer's Reference 2550

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals