4 System
4.2 I/O Config
4.2.8 USB Connection Status (Remote Query Only)
Lets you determine the current status of the USB connection.
Remote
Command
:SYSTem:COMMunicate:USB:STATus?
Example
:SYST:COMM:USB:STAT?
Notes SUSPended Indicates that the USB bus is currently in its suspended state. The bus is in the
suspended state when:
The bus is not connected to any controller
The controller is currently powered off
The controller has explicitly placed the USB device into the suspended state
When in the suspended state, no USB activity, including start of frame packets are received
ACTive Indicates that the USB device is in the active state. When the device is in the active state, it
receives periodic frame starts, but is not necessarily receiving or transmitting data
State Saved No
Range
SUSPended | ACTive
4.2.9 USB Packet Count (Remote Query Only)
Lets you determine the number of packets received and transmitted on the USB
bus.
Remote
Command
:SYSTem:COMMunicate:USB:PACKets?
Example
:SYST:COMM:USB:PACK?
Notes Two integers are returned:
1.
The number of packets received since application invocation
2.
The number of packets transmitted since application invocation
If no packets have been received or transmitted, the response is 0,0
The packet count is initialized to 0,0 when the instrument application is started
State Saved No
Vector Modulation Analyzer Mode User's &Programmer's Reference 1856

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals