4 System
4.2 I/O Config
Disconnecting the individual interface and session releases the lock if the lock is granted to the
interface and session
A Device Clear over any interface and session releases the lock, regardless of the interface and session
which obtained the lock
The following queries are permitted over any interface and session, even if an interface has the
instrument locked:
*IDN?
*OPT?
*STB?
*ESR?
:SYSTem:DATE?
:SYSTem:TIME?
:SYSTem:PON:TIME?
Queries in the :STATus subsystem
Queries in the :SYSTem:ERRor subsystem
Queries in the :SYSTem:LKEY subsystem
Queries in the :SYSTem:LOCK subsystem
Queries in the :SYSTem:METRics subsystem
Queries in the :SYSTem:MODule subsystem
All other commands and queries result in error: -203,”Command protected; Instrument locked by
another I/O session”
State Saved Not part of Save/Recall
4.2.10.2 Unlock Remote I/O Session (Remote Command only)
You can unlock the SCPI control of the current I/O Interface and Session by sending
:SYSTem:LOCK:RELease. Lock requests on an individual interface and session can
be nested, and each request increases an internal lock count by 1. For every granted
request, you will need to perform a release. The lock is not relinquished until the
internal lock count reaches 0.
Remote
Command
:SYSTem:LOCK:RELease
Example
:SYST:LOCK:REL
Notes When the instrument is unlocked, Bit 0 is cleared in the Operation Instrument status register
1859 Vector Modulation Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals