3 VMA Mode
3.4 IQ Waveform Measurement
If the selected marker is OFF, it is turned ON before the minimum search is
performed.
Remote Command
:CALCulate:WAVeform:MARKer[1]|2|…|12:MINimum
Example
:CALC:WAV:MARK:MIN
Notes Sending this command selects the subopcoded marker
State Saved Not part of saved state
Marker Delta
Pressing this control has the same effect as pressing Delta in "Marker Mode" on
page 747 on the Settings tab. The selected marker becomes a Delta marker. If the
selected marker is already a Delta marker, the reference marker is moved to the
current position of the selected marker, thus resetting the delta to zero.
The control is duplicated here to allow you to conveniently perform a Peak Search
and change the marker’s control mode to Delta, without having to access two
separate menus.
3.4.7.4 Pk Search Config
Contains controls that let you set up the Peak Search functions.
Since the Pk Search Config functions are independent of the selected Marker, the
Select Marker control does not appear while in Pk Search Config.
Peak Search Range
Lets you specify the range of trace to perform peak search actions specified in the
Peak Search menu.
Option SCPI Behavior
Full
FULL
Peak Search actions will be performed on the entire trace
Manual
MANual
Specifies the range of the trace to which Peak Search actions will
apply
Remote
Command
:CALCulate:WAVeform:MARKer:PEAK:SEARch:RANGe FULL | MANual
:CALCulate:WAVeform:MARKer:PEAK:SEARch:RANGe FULL | MANual
Example
:CALC:WAV:MARK:PEAK:SEAR:RANG FULL
Preset
FULL
State Saved Saved in instrument state
Range
FULL | MANual
751 Vector Modulation Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals