EasyManuals Logo

Keysight x-series Programmer's Reference

Keysight x-series
2650 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #184 background imageLoading...
Page #184 background image
3 VMA Mode
3.1 Digital Demod Measurement
For this measurement, the value of n is derived by concatenating two 2-digit sub-
fields: Segment Index (SI) and Result Index (RI).
For details, see "Index Sub-Fields" on page 184.
For an overview of all available results, see "Summary of Result Types" on page 185.
In cases where results depend on Modulation Format, details have been moved to
separate tables for clarity.
3.1.1 Index Sub-Fields
The general formula for calculating the desired value of n is:
n = 100 * SI + RI
Sub-field Variable Definition
Most Significant 2 Digits
SI
Segment Index
00 Segment1
General results not specific to a segment
01 Segment2
02 Segment3
03 Segment4
04 Segment5
05 Segment6
Least Significant 2 Digits
RI
Result Index
The information returned is for the specified SI:
00 Captured IQ data
01 Metrics
02 Limit test results
03 IQ data for Search Time
04 IQ data for Time (Meas Interval)
05 Spectrum Trace
06 IQ Meas Time
07 IQ Ref Time
08 IQ Meas Spectrum
09 IQ Ref Spectrum
10 Error Vector Time
11 EVM Spectrum
12 Magnitude Error
Vector Modulation Analyzer Mode User's &Programmer's Reference 184

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)