3 VMA Mode
3.2 Custom OFDM Measurement
Remote Command
:CALCulate:OFDM:MARKer[1]|2|…|12[:SET]:CENTer
Example
:CALC:OFDM:MARK4:CENT
Notes Sending this command selects the subopcoded marker
If specified marker is Off, this command turns it on at the center of the screen as a Normal type marker
Dependencies Not available (grayed-out) when x-axis is the time domain
Couplings All the usual couplings associated with setting Center Frequency apply
3.2.8 Meas Setup
Contains functions for setting up the measurement parameters, and also contains
functions for setting up parameters global to all measurements in the Mode.
3.2.8.1 Component Carrier
Lists the available component carriers.
3.2.8.2 Settings
Lets you set measurement parameters.
Avg|Hold Number
Specifies the number of N averages that will be used for the measurement. After the
specified number (average counts) have been averaged, the averaging mode
(termination control) setting determines the averaging action.
Remote Command
[:SENSe]:OFDM:AVERage:COUNt <integer>
[:SENSe]:OFDM:AVERage:COUNt?
Example
:OFDM:AVER:COUN 1000
:OFDM:AVER:COUN?
Preset 10
State Saved Yes
Min 1
Max 10000
Averaging On/Off
Turns averaging on or off.
489 Vector Modulation Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals