9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Bit Condition Operation
10 Reference missing
or out of range
The signal being fed to a reference input is missing or too high or
low in frequency for the reference to lock
Filter Registers
"Questionable Frequency Condition" on page 2602
"Questionable Frequency Enable" on page 2602
"Questionable Frequency Event Query" on page 2603
"Questionable Frequency Negative Transition" on page 2603
"Questionable Frequency Positive Transition" on page 2604
Questionable Frequency Condition
Returns the decimal value of the sum of the bits in the Questionable Frequency
Condition register.
NOTE
The data in this register is continuously updated and reflects the current
conditions.
Remote Command
:STATus:QUEStionable:FREQuency:CONDition?
Example
:STAT:QUES:FREQ:COND?
Preset 0
Status Bits/OPC depend-
encies
Sequential command
Questionable Frequency Enable
Determines which bits in the "Questionable Frequency Condition" on page 2602
Register will set bits in the "Questionable Temperature Event Query" on page 2600
register, which also sets the Frequency Summary bit (bit 5) in the "Questionable
Register" on page 2592.
The variable <integer> is the sum of the decimal values of the bits you want to
enable.
Remote
Command
:STATus:QUEStionable:FREQuency:ENABle <integer>
:STATus:QUEStionable:FREQuency:ENABle?
Example
:STAT:QUES:FREQ:ENAB 2
Frequency Reference Unlocked will be reported to the Frequency Summary of the Status
Vector Modulation Analyzer Mode User's &Programmer's Reference 2602

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals