Page #632 background image
3 VMA Mode
3.3 Monitor Spectrum Measurement
Remote
Command
:CALCulate:MONitor:MARKer[1]|2|...|12:MODE POSition | DELTa | OFF
:CALCulate:MONitor:MARKer[1]|2|...|12:MODE?
Example
:CALC:MON:MARK:MODE POS
:CALC:MON:MARK:MODE?
Preset
OFF
State Saved Saved in instrument state
Range
POSition|DELTa|OFF
Annotation Mkr # <X value> and <Marker value> upper right on graph
Backward Compatibility SCPI Commands
Sets or queries the state of a marker. Setting a marker that is OFF to ON (1) puts it in
Normal mode and places it at the center of the screen.
Remote
Command
:CALCulate:MONitor:MARKer[1]|2|...|12:STATe OFF | ON | 0 | 1
:CALCulate:MONitor:MARKer[1]|2|...|12:STATe?
Example
:CALC:MON:MARK3:STAT ON
:CALC:MON:MARK3:STAT?
Preset
OFF
State Saved Saved in instrument state
Range
OFF|ON
Delta Marker (Reset Delta)
Pressing this button has exactly the same effect as selecting the Delta selection in
"Marker Mode" on page 631. The selected marker becomes a Delta Marker. If the
selected marker is already a Delta marker, the reference marker is moved to the
current position of the selected marker, thus resetting the Delta to zero.
Marker Settings Diagram
Lets you configure the Marker system using a visual utility.
All Markers Off
Turns off all markers.
Remote Command
:CALCulate:MONitor:MARKer:AOFF
Example
:CALC:MON:MARK:AOFF
Vector Modulation Analyzer Mode User's &Programmer's Reference 632

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals