3 VMA Mode
3.4 IQ Waveform Measurement
Preset 0.2
State Saved Saved in instrument state
Min 0.01
Max 1.00
Backwards Compatibility
SCPI
[:SENSe]:WAVeform:WBIF:FILTer:ALPHa
Channel Filter Bandwidth (Backwards Compatibility Remote Command Only)
Dependencies Only available when Option DP2, B40, or wider IF Bandwidth option is installed
Couplings The value is determined by the following equation
ChannelFilterBwBwcc = (ChannelFilterBw/(DigitalIFBw*OverSampleRatio))
Preset 0.8
State Saved Saved in instrument state
Min 0.01
Max 1.0
Backwards
Compatibility SCPI
[:SENSe]:WAVeform:WBIF:FILTer:BANDwidth <real>
[:SENSe]:WAVeform:WBIF:FILTer:BANDwidth?
3.4.5 Display
Lets you configure display items for the current Mode, Measurement, View, or
Window.
3.4.5.1 View
Contains controls for selecting the current View, and for editing User Views.
View
See "Views" on page 674.
User View
Lets you choose a View from the saved User Views for the current measurement.
This panel only appears if a User View exists for the current measurement.
Remote
:DISPlay:VIEW:ADVanced:SELect <alphanumeric>
Vector Modulation Analyzer Mode User's &Programmer's Reference 730

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)